Unitron TrueFit software instruction guide v2.2
4
• Save: saves the current fitting information. Choose from Save to hearing
instrument, Save to database, or Save to accessories (if accessory added
to fitting).
• Close Unitron TrueFit: exits the software application. You will be prompted to save
your fitting information if you have not done so before closing Unitron TrueFit .
Hearing Instrument
• Save: saves the current fitting information. Choose from Save to hearing
instrument, Save to database, or Save to accessories (if accessory added to
fitting).
• Test Mode: sets the hearing instrument into a test mode so you can compare the
electroacoustic performance of the device to the published ANSI / IEC
specification data sheets.
• Verification Mode: sets the hearing instrument into a verification mode to allow
testing without the interference of adaptive features. There are 3 options:
• Adaptive features active: default state.
• Real Ear Verification: all adaptive features are disabled except for Natural Sound
Balance and input selection.
• 2cc Coupler Verification: all adaptive features and input selection are disabled.
• HI Update: checks if newer functionality is available for the connected hearing
instrument. If an update is available, the instrument maintains its fine tuning and
new features become enabled.
Note: Quantum Pro and Moxi Pro devices cannot be updated using iCube. Only
NOAHlink and HI-PRO can be used.
• HI Repair / Reset: opens a wizard to reset the firmware of a hearing instrument.
Only use this when directed by an error message. If this was not successful, please
try again before sending the hearing instrument back to your distributor. The client
record stored in the database will not be affected.
• Transfer Fitting: provides instructions on how to transfer a fitting to or from
another hearing instrument.
Options
• Preferences: generates the user preferences dialog box with the following
options:
General:
• Language: select the preferred language for text and sound files.
• Presentation: select your preferred level of animations.
• Support: generate Unitron TrueFit logging information for support.
• Flex: enter code here to enable Flex:upgrade functionality. Choose the
appropriate trial period defaults for Flex:trial and Flex:upgrade.
Fitting Session:
• Diagnostics: choose the preferred bone conduction symbol and view/adjust the
DSL v5 ABR nHL to eHL correction values.
• Fitting: indicate if the speech verification graph option will be available from the
curve display options on the Pre Fitting screen. Choose whether to enable
password protection to keep fitting information confidential. Following
activation of this option, the fitter can insert their desired password into the