Unitron TrueFit software instruction guide v2.2
12
You can create a blank session for a client by clicking on the client information and
then clicking Blank Session icon.
Experience Level
Select the client’s experience level with hearing
instruments. Choose from:
i. First Time
ii. Short Term / Part Time
iii. Long Term
Note: The selected experience level with hearing
instruments will impact the Quick Fit for the Automatic
Adaptation Manager.
Note: Be sure to Quick Fit after changing the
Experience Level.
Audiogram
When running Unitron TrueFit under NOAH, the
audiogram screen is primarily for reference. The
transducer used for audiometry and the tinnitus
match sections are used as described below. To enter
or change all other audiometric information, use the
NOAH audiogram module.
When running the Unitron TrueFit standalone version,
the audiogram screen provides the ability to enter the
client’s audiometric test results.
Select the transducer used for audiometry from the
dropdown menu. Choose from: Headphones, Insert
earphones, Custom earpiece (earmold); Loudspeaker
(0, 45, or 90 degrees azimuth); ABR insert earphones
(eHL or nHL); ABR custom earmold (eHL or nHL).
Note: ABR eHL and ABR nHL data is only used with the
DSL v5 fitting formula.
To complete the audiogram, enter the data for the appropriate ear by clicking on the
graph. Right clicking the graph provides the option to: Set point, Remove point, Remove
all points or Copy (all curves) or Copy selected curve to the other ear. In addition, right
clicking provides the option of specifying the measurement type for each data point;
choose from: Unmasked, No response, Unmeasurable, Masked, Masked and no
response or Masked and unmeasurable.
To enter air conduction data, click on the AC button.
To enter bone conduction data, click on the BC button.
To enter uncomfortable loudness level data, click on the UCL button.
Note: NAL-NL1 and NAL-NL2 do not use UCL data for target generation.
To copy the entered curve from one ear to the next, click on the arrow icons above the
graph for the respective ear.
To delete all of the curves on the audiogram, click on the Remove (Delete)icon.
Click on Use Audiogram to use the Audiogram for fitting formula targets.
Click on Use In Situ to use the measured In Situ data for fitting formula targets.